Specifications |
Model |
Orion 3-Star Plus Conductivity Benchtop Meter |
Conductivity Mode |
Range |
0.000 to 3000mS/cm, auto resolution with cell constant dependence |
Resolution |
4 significant digits down to 0.001µS/cm, dependent on cell constant |
Relative Accuracy |
±0.5% ±1 digit or 0.01µS/cm, whichever is greater |
Cell Constants |
0.001 to 199.9cm-1 |
Cell Compatibility |
2 Cell, 4 Cell, or Planar |
Calibration |
DirectCal*/AUTO-CAL*: 1 to 5 points/1 to 3 points |
Reference Temperature |
5°C, 10°C,15°C, 20°C or 25°C (default); autoranging |
Resistivity Mode |
Range |
0.0001 to 100megohm |
Resolution |
Autoranging |
Relative Accuracy |
±0.5% ±1 digit |
Salinity Mode |
Range |
0.1 to 80.0ppt NaCl equivalent; 0.1 to 42ppt practical |
Resolution |
0.1ppt |
Relative Accuracy |
±0.1ppt ±1 digit |
TDS Mode |
Range |
0 to 19,999mg/L |
Resolution |
1mg/L |
Relative Accuracy |
±0.5% ±1 digit |
Adjustable TDS Factor |
0.0 to 10.0% |
Temperature Mode |
Range |
-5° to +105°C |
Resolution |
0.1°C up to 99.9°C, 1.0 over 99.9°C |
Relative Accuracy |
±0.1°, ±1 digit |
Temperature Compensation |
nLF, linear |
Display |
Custom LCD; simultaneous display of selected measurement and temperature, and current electrode condition |
Inputs/Outputs |
Inputs |
8-Pin Mini-DIN Locking for Conductivity; Phono jack for stirrer |
Sensor Types |
2 Cell, 4 Cell, or Planar Conductivity |
Outputs |
Bidirectional RS-232, stereo jack, optional USB inteface cable |
Data Management |
GLP Electrode Veri?cationSaves 1000 points by time, stability or manual read with time/date stampCalibration log of last 10 calibrationsUser method storage for 10 proceduresAdvanced self-test and diagnostics routinesPassword protectionDownloadable software and methods updates via Internet |
Regulatory and Safety Compliance |
CE, CSA, UL, TUV, FCC Class A limits |
Environmental Operating Conditions |
Ambient Operating Temperature |
5° to +45°C |
Relative Humidity |
5 to 85% noncondensing |
IP Rating |
IP54 compliance: splashproof and dustproof |
Stability |
SMART STABILITY* and SMART AVERAGING*: Resolution dependent/automatic |
Power Supply |
Universal AC power adapter or 4 x 1.5V AA (optional) |
Warranty |